Applications
Biological safety cabinets
are the primary barrier against exposure to aerosols that may be produced
from common microbiology procedures. BSC2000 Series cabinets provide
personnel and environment protection in applications involving
biological agents which are classified as being of ordinary, potential
or special hazard.
These cabinets are not designed
for use with biological agents of extreme hazard. For microbiology applications
requiring product protection as well as personnel and environment protection,
BH2000
Class II cabinets should be specified.
Australian and International
Standards Compliance
Please refer to our Standards
page.
Description
BSC2000 Series cabinets are
self-contained 100% exhaust containment enclosures which operate independently
of other air-handling systems. Air from the work zone passes through a
two-stage, high-efficiency filter system before being exhausted to the
room. Cabinets are produced with work zone width of 90cm, 120cm or 180cm
and may be installed on standard laboratory benches or on optional floor
stands.
Personnel and environment
protection is provided by an inward flow of room air at the work opening
and high-efficiency particulate air (HEPA) filtration of exhaust air. The
inward airflow sweeps aerosols generated during the work process away from
the operator into the filter system. A prefilter protects the HEPA filter
from large particles, and the HEPA filter removes particulate contaminants
down to sub-micron size from the exhaust air.
All positive pressure zones
and filter seals are surrounded by negative pressure zones, so as to contain
potentially hazardous aerosols. BSCA90 and BSCA120 cabinets have exhaust
discharge on the right hand side, and optional left hand side or top exhaust
is available. BSCA180 cabinets are produced with top exhaust only.
Embodying advanced system-monitoring
technology, enhanced ergonomics and styling, BSC2000 Series cabinets set
a new standard in safety cabinet design and performance.
Construction
Cabinet.
Constructed in electro-galvanised
steel with joints welded using a gas-shielded arc process. This method
produces a robust, gas-fight housing which is able to withstand the rigours
of transport and handling. Exterior finish is in a special baked enamel
which has been developed for laboratory equipment.
Work zone.
Constructed in Grade 304
stainless steel with 2B finish. Corners are radiused and crevice-free for
ease of cleaning and all surfaces are carefully dressed to remove sharp
edges.
A pneumatic mechanism opens
and closes the viewing window without the need for external fasteners or
catches. The window is self-supporting in the fully-open position to facilitate
cleaning and access for large items. Opening the window with the cabinet
running automatically engages a boost mode for enhanced containment by
activating maximum exhaust airflow and the alarm systems.
Fans.
A direct drive fan is fitted
with a speed controller to enable airflows to be maintained through filter
life. Audible and visible alarms with rechargeable
battery back-up signify any variation in airflow. A manometer gauge located
in the work zone monitors fan/filter system pressures.
HEPA filters.
Clyde-Apac HEPA filters
are manufactured to the most exacting standards for critical applications.
Each filter is individually certified to achieve an arrestance efficiency
of not less than 99.995% to the stringent BS3928 Sodium Flame test. All
testing is conducted in a NATA-registered factory laboratory and a NATA-endorsed
test label, being an extract of the test report, is affixed to each filter.
Physical Data
|
Model
|
Exhaust
Discharge
|
Overall dimensions
(mm)
|
Work zone dimensions
(mm)
|
Weight
(kg)
|
|
|
|
W
|
D
|
H
|
W
|
D
|
H
|
|
|
BSCA 90
|
LHS or RHS
|
1050
|
765
|
1305
|
870
|
560
|
710
|
165
|
|
BSCA 90
|
Top
|
1050
|
765
|
1530
|
870
|
560
|
710
|
165
|
|
BSCA 120
|
LHS or RHS
|
1365
|
765
|
1305
|
1180
|
560
|
710
|
190
|
|
BSCA 120
|
Top
|
1340
|
765
|
1530
|
1180
|
560
|
710
|
190
|
|
BSCA 180
|
Top
|
1950
|
765
|
1530
|
1790
|
560
|
710
|
285
|
Electrical and diagnostics.
A low-voltage touch-control
panel is located on the front of the cabinet. A microprocessor-controlled
diagnostic system provides continuous monitoring of critical cabinet functions
with a digital display indicating the nature of any malfunction. The boost
mode and a post-use over-run mode may be manually selected from the control
panel.
An integral fluorescent lamp
housing reduces heat build-up near the operator. Glare-free lamps provide
a minimum lighting intensity of 800 lux at the work floor. A splashproof,
general purpose power outlet is fitted in the work zone.
Quality Assurance and
Control
Clyde-Apac laminar flow
and HEPA filter products are manufactured in Australia under an accredited
ISO-9001 Quality Management system. Please refer to our Standards
page.
Standard features
-
Low voltage touch controls
-
Diagnostics system with digital
display monitoring critical functions
-
Alarms and Boost mode
automatically engaged when viewing window is open
-
Boost mode selectable at control
panel
-
Selectable post-use over-run
timer
-
Pneumatically-operated viewing
window
-
Hourmeter to record operating
time
-
Magnehelic gauge to monitor
fan systems
-
Splashproof power outlet in
work zone
-
Stainless steel work opening
cover
-
Comprehensive operation and
maintenance manual
Options
-
Additional power outlet
-
Germicidal UV lamp
-
Gas tap with solenoid valve
-
Vacuum tap
-
Floor stand
-
Exhaust discharge on left hand
side or top for BSCA120
-
Fumigation adaptor panels for
work opening and exhaust
On-site testing
BSC2000 Series cabinets
are factory-tested and certified by a NATA-registered laboratory. Additional
testing and certification
should be carried out on site prior to use, and subsequently:
-
After any electrical or mechanical
maintenance
-
After filter replacement
-
After re-location
-
At least annually
-
In special circumstances, such
as a significant change in the work programme, or where unsafe cabinet
operation is suspected
Laminar
Air Flow Services, a NATA-registered laboratory and a division of Clyde-Apac,
provides comprehensive on-site maintenance, testing and certification services
for safety cabinets, laminar flow work stations, cleanrooms
and HEPA filter installations. This service is available from
fully-equipped, NATA-registered laboratories in major Australian and International
centres.
Due to ongoing research
& development, Clyde-Apac reserves the right to modify the product
& specifications without notice.
|