LFC 60 - Laboratory fume cabinet
Clyde Apac

 
The LFC60 fume cabinet is a compact, recirculating fume cabinet that has been designed to provide operator safety and ease of use in laboratory applications. Capture of fumes is achieved by inward airflow at a velocity of > 0.5 m/s and carbon filtration of exhaust air. The upper housing contains the mains switch, low-airflow alarm, speed control and carbon filter. 

Cabinets are supplied with an open base and can be positioned over an existing utility such as a sink or be placed directly on to a flat bench.An optional work tray, manufactured from polypropylene, is available if containment of spills is required.

The standard activated-carbon filter filter* is suitable for use with a wide range of common laboratory fumes.
 
 
 
 
 
 
 
 
 



Specifications.
 
Construction Baked enamel coated top module. Airflow m3/h 200
Colour Grey Face velocity m/s > 0.5
Windows Clear Perspex Face velocity m/s 15 W
Spillage tray Optional polypropylene Carbon filter No. & mass 1 x 4 kg
Fan Centrifugal Noise level dB(A) 58
Electrical Supply 240V, 50Hz
Controls Mains power On/Off
Monitoring Alarm for airflow and filter condition

Dimensions
 
Width 600mm
Depth 580mm
Overall height 800mm
Working height 500mm
Work area 580 x 520

*Special-purpose carbon filters for specific applications are available to order.


Laminar Air Flow Services, a NATA-registered laboratory and a division of Clyde-Apac, provides comprehensive on-site maintenance, testing and certification services for safety cabinets, laminar flow work  stations,  cleanrooms  and  HEPA  filter installations. This service is available from fully- equipped laboratories in major Australian centres. 
Due to ongoing research & development, Clyde-Apac reserves the right to modify the product & specifications without notice.
Top of Page